Automatic melt flow rate tester

Function summary

       Automatic melt flow rate tester (also known as melt index meter) is a special instrument for measuring the melt flow rate of thermoplastics under certain conditions。

       The melt flow rate (melt index) of thermoplastics refers to the mass or melt volume of the melt through the standard die every 10 minutes at a certain temperature and load, expressed by the MFR (MI) or MVR value。The test is mainly used for the quality inspection and storage inspection of raw materials and products of thermoplastics and chemical fibers。

       Machine height automation,Integrated press, lifting weights and cleaning processes;Through the clasp on the device,Can accurately control each pressing process;The equipment is in the process of testing,The loading and lifting of weights are completed automatically without manual intervention;After the test is completed, it is equipped with special cleaning components,Barrel cleaning is done at the touch of a button。


Relevant standard

  • GB/T 3682-2018 Determination of melt mass flow rate and melt volume flow rate of thermoplastics

  • JB/T5456-2016 Melt flow rate instrument technical conditions

  • ISO 1133-2011 Polymer materials - Mass and volume flow rate testing

  • JJG878-1994 (2005) Verification regulation of melt flow rate instrument


Technical index

Specification and model

MFI452B-A

Measurement method

Mass method, volume method

Temperature control range

50℃~450℃

Temperature control controls fluctuations

± 0.5℃

Temperature control 4h drift

≤±0.5℃

Temperature Controls the temperature gradient

≤0.5℃ (the upper end of the inner die of the barrel is 10 ~ 70mm in the tropical zone)

Temperature control resolution

0.1℃

Cylinder temperature recovery time after feeding

≤4min

Time clock range

0s~6000s

Timing clock resolution

0.01s

Displacement error

≤±0.02mm

Displacement resolution

0.01mm

Automatic cutting device

Set the time interval (adjustable from 2s to 2000s) to automatically complete the cutting

Inside diameter of the die

Φ2.095± 0.005mm

Inside diameter of barrel

Φ9.550±0.007mm

Outside diameter of piston head

Φ9.474±0.007mm

Weight load accuracy

≤±0.5%

Weight load

0.325kg、1.20kg、2.16kg、3.80kg、5.00kg、10.00kg、21.60kg

Measuring range

0.1 ~ 1500g/10min

Recommended assay

When MFR is 0.04 ~ 30g/10min, suitable for measuring flow speed by mass method;

当MFR> 30g/10min时,适合用体积法测流速;

National standard sample (PE) sample repetition accuracy

≤2%

National standard sample (PE) sample accuracy

≤5%

Rated service pressure

0.40Mpa

Main engine dimensions

630mm×550mm×1200mm

Operating voltage

AC220±10% V

Rated power

1.5kW

重量